© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
1-10
Figure 1-6
Probe Compensation
2
)
Set the probe attenuation of the oscilloscope to X10. (The default is X1).
Method of set the probe attenuation:
Right click the
button into the Channel menu. In the Probe menu, select the
probe attenuation. Now select the
1:10
option.
Figure 1-7
Figure1-8
To Set the Probe To Set the Probe Attenuation
3
)
Attach the tip of probe and ground nip to the Connector of Probe compensator.
Click the
button. A square wave will be displayed within a few seconds.
(Approximately 1 kHz, 3 V, peak- to- peak).
4
)
Inspect Then CH2 with the same method. Click the
button to turn Channel 1
off, click the
button to turn Channel 2 on. Repeat steps 2 and 3.
Probe
attenuation
Probe
compensator