RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-69
1)
Select the
Manual Cursor
mode for cursor measurement; click the buttons as:
Analyze
Æ
Cursor
Æ
Manual Cursor.
2)
Select channel
Source
for measurements, click the buttons as:
Source
Æ
CH1, CH2, MATH (1, 2, 3 and 4) or LA (for VS5000D only).
3)
Select the Cursor type by clicking the appropriate buttons; in the Manual Cursor
window, there are X and Y operation windows to select the cursor type.
4)
Move the cursors to adjust the increment between the cursors: (Details in the
following Table)
Table 2-31 Definition of the Manual Cursors
Increment Cursor
Operation
X
Cursor A
Cursor B
Click the
button to adjust the
cursor A and B horizontally.
Y
Cursor A
Cursor B
Click the
button to adjust the
cursor A and B vertically.
NOTE:
Cursor could be moved only when the cursor function menu is displayed.
5)
To get measurement values:
z
Position of Cursor 1 (time cursor centered on the midpoint of screen; Voltage
cursor centered on channel ground level).
z
Position of Cursor 2 (same as above).
z
Horizontal space between cursor 1 and 2 (
△
X): Time between cursors (1/
△
X),
units in Hz, kHz, MHz, GHz.
z
Vertical space between cursor 1 and 2 (
△
Y): Voltage between cursors.
If the source is LA (for VS5000D only), the measurement value is as follows:
z
Position of Cursor 1 (time cursor centered on the midpoint of screen).
z
Position of Cursor 2 (time cursor centered on the midpoint of screen).
z
Cursor 1 value: Hex.
z
Cursor 1 value: Binary.
z
Cursor 2 value: Hex.
z
Cursor 2 value: Binary.
NOTE:
The values will be automatically displayed on the right upper corner of screen
when the cursor function menu is hidden or displaying other menus.
Содержание VS5000 Series
Страница 2: ......
Страница 37: ...RIGOL 2008 RIGOL Technologies Inc User s Guide for VS5000 Series 2 5 Figure 2 3 DC Coupling Setting DC Coupling...
Страница 70: ...2008 RIGOL Technologies Inc User s Guide for VS5000 Series 2 38 Figure 2 24 Video trigger Line Synchronization...
Страница 83: ...RIGOL 2008 RIGOL Technologies Inc User s Guide for VS5000 Series 2 51 Figure 2 31 Peak Detect...
Страница 114: ......
Страница 118: ......
Страница 129: ...RIGOL 2008 RIGOL Technologies Inc User s Guide for VS5000 Series 3 11 Figure 3 10 FFT Measurement Cursor Type X...
Страница 131: ...RIGOL 2008 RIGOL Technologies Inc User s Guide for VS5000 Series 3 13 Figure 3 11 Pass Fail Test...