Chapter 2 Command Systems
RIGOL
Programming Guide for DSA1000A series
2-85
results, such as
:CHP:AVER:COUN 100
The query returns “100”.
Explanations
<integer> ranges from 1 to 1000.
Default
10
Front Panel
Meas, Chan Pwr
Meas Setup, Avg Num
51.
[:SENSe]:CHPower:AVERage[:STATe]
Syntax
[:SENSe]:CHPower:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:CHPower:AVERage[:STATe]?
Function
Turns on or off the average function of the channel power
measurement.
The query returns “0” or “1”.
Default
OFF|0
Front Panel
Meas, Chan Pwr
Meas Setup, Avg Num, On/Off
52.
[:SENSe]:CHPower:AVERage:TCONtrol
Syntax
[:SENSe]:CHPower:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:CHPower:AVERage:TCONtrol?
Function
Sets the average mode of the channel power measurement.
The query returns “EXP” or “REP”.
Explanations
In Exp mode, the result is from the exponent average calculation
of the current N (specified by [:SENSe]:CHPower:AVERage:COUNt
measurements.
In REP mode, the result is the arithmetic average of the current N
(specified by [:SENSe]:CHPower:AVERage:COUNt)
measurements.
Default
EXPonential
Front Panel
Meas, Chan Pwr
Meas Setup, Avg Mode, Exp/Repeat
53.
[:SENSe]:CHPower:BANDwidth:INTegration
Syntax
[:SENSe]:CHPower:BANDwidth:INTegration <freq>
[:SENSe]:CHPower:BANDwidth:INTegration?
Function
Sets the integration bandwidth for the channel power measurement,
Содержание DSA1000A Series
Страница 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Страница 2: ......
Страница 14: ......
Страница 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Страница 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Страница 155: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 21 7 Execute results...
Страница 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...