Chapter 2 Command Systems
RIGOL
Programming Guide for DSA1000A series
2-91
measurements.
In REP mode, the result is the arithmetic average of the current N
(specified by [:SENSe]:HDISt:AVERage:COUNt) measurements.
Default
EXPonential
Front Panel
Meas, Harmo Dist
Meas Setup, Avg Mode, Exp/Repeat
70.
[:SENSe]:HDISt:NUMBers
Syntax
[:SENSe]:HDISt:NUMBers <integer>
[:SENSe]:HDISt:NUMBers?
Function
Sets the number of harmonics for carrier measurement, such as
:HDIS:NUMB 5
The query returns “5”.
Explanations
<integer> ranges from 2 to 10.
Default
10
Front Panel
Meas, Harmo Dist
Meas Setup, NO.of Harmo
71.
[:SENSe]:HDISt:TIME
Syntax
[:SENSe]:HDISt:TIME <time>
[:SENSe]:HDISt:TIME?
Function
Sets the sweep time of the Harmonic measurement, the default unit is
s, such as
HDIS:TIME 0.01
The query returns “1.000000E-02”.
Explanations
<time> ranges from 20 us to 3 ks.
Default
10 ms
Front Panel
Meas, Harmo Dist
Meas Setup, Sweep Time
72.
[:SENSe]:HDISt:TIME:AUTO[:STATe]
Syntax
[:SENSe]:HDISt:TIME:AUTO[:STATe] OFF|ON|0|1
[:SENSe]:HDISt:TIME:AUTO[:STATe]?
Function
Turns on or off the Auto function of sweep time for the harmonic
measurement.
The query returns “0” or “1”.
Default
ON|1
Содержание DSA1000A Series
Страница 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Страница 2: ......
Страница 14: ......
Страница 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Страница 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Страница 155: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 21 7 Execute results...
Страница 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...