RIGOL
Chapter 2 Command Systems
Programming Guide for DSA1000A series
2-66
power.
Explanations
The power unit of returned value is the same as the unit of
Y-axis.
The command transmits the data from the latest measurement
to the ouput buffer.
Return Value Such as -1.01
12.
:READ:EBWidth?
Syntax
:READ:EBWidth?
Function
The command executes a measurement and returns the
measurement result of emission bandwidth in Hz.
Explanations The command transmits the data from the latest measurement to
the ouput buffer.
Return Value Such as 50000
13.
:READ:HARMonics:AMPLitude:ALL?
Syntax
:READ:HARMonics:AMPLitude:ALL?
Function
The command executes a measurement and returns the amplitude
value of the first ten harmonics at comma intervals and having the
same unit as the current amplitude.
Explanations
If there are less than ten harmonics to measrure, the query
returns nothing for those unmeasured harmonics.
The unit of returned value is the same as the unit of Y-axis.
The command transmits the data from the latest measurement
to the ouput buffer.
Return Value Such as
-1.01,-6.01,-7.01,-7.01,
-7.01,-7.01,-7.01,
-7.01,
-7.01,-7.01
14.
:READ:HARMonics:AMPLitude? <n>
Syntax
:READ:HARMonics:AMPLitude? <n>
Function
The command executes a measurement and returns the specified
harmonic amplitude, such as
:READ:HARM:AMPL? 1
Explanations
<n> is an integer wthin 1 and 10.
The unit of returned value is the same as the unit of Y-axis.
Содержание DSA1000A Series
Страница 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Страница 2: ......
Страница 14: ......
Страница 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Страница 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Страница 155: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 21 7 Execute results...
Страница 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...