Chapter 2 Command Systems
RIGOL
Programming Guide for DSA1000A series
2-69
values at comma intervals:
Frequency of Base Lower (Hz), Amplitude
Frequency of Base Upper (Hz), Amplitude
Frequency of 3rd Order Lower (Hz), Amplitude and Intercept of TOI
Frequency of 3rd Order Upper (Hz), Amplitude and Intercept of TOI
Explanations
The amplitude unit of returned value is the same as the unit of
Y-axis.
The command transmits the data from the latest measurement
to the ouput buffer.
Return Value
Such as
1500450000,-8.01,
1500450000,-8.01,
1500450000,-8.01,-8.01,
1500450000,-8.01,-8.01
23.
:READ:TOIntercept:IP3?
Syntax
:READ:TOIntercept:IP3?
Function
The command executes a measurement and returns the smaller of
the TOI intercept between the 3rd Order Lower and the 3rd Order
Upper.
Explanations The command transmits the data from the latest measurement to
the ouput buffer.
Return Value
Such as -8.01
24.
:READ:TPOWer?
Syntax
:READ:TPOWer?
Function
The command executes a measurement and returns the
Time-domain power measurement result.
Explanations
The power unit of returned value is the same as the unit of
Y-axis.
The command transmits the data from the latest measurement
to the ouput buffer.
Return Value
Such as -1.01
Содержание DSA1000A Series
Страница 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Страница 2: ......
Страница 14: ......
Страница 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Страница 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Страница 155: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 21 7 Execute results...
Страница 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...