Chapter 2 Command Systems
RIGOL
Programming Guide for DSA1000A series
2-77
Explanations
<time> ranges from 5 ms to 1000000 s.
Default
100 ms
Front Panel
Demod, Demod, AM/FM, Demod Setup, Demod Time
17.
[:SENSe]:DETector[:FUNCtion]
Syntax
[:SENSe]:DETector[:FUNCtion]
NEGative|NORMal|POSitive|RMS|SAMPle|VAVerage
[:SENSe]:DETector[:FUNCtion]?
Function
Sets the detection type.
The query returns “NEG”, “NORM”, “POS”, “RMS”, “SAMP” or “VAV”.
Default
POSitive
Front Panel
Detector, Pos Peak/Neg Peak/Sample/Normal/RMS Avg/Voltage Avg
18.
[:SENSe]:FREQuency:CENTer
Syntax
[:SENSe]:FREQuency:CENTer <freq>
[:SENSe]:FREQuency:CENTer?
Function
Sets the center frequency, the default unit is Hz, such as
:FREQ:CENT 1000000 or :FREQ:CENT 1MHZ
The query returns “1000000”.
Explanations
<freq> ranges from 0 Hz to 3 GHz.
Default
1.5 GHz
Front Panel
FREQ, Center Freq
19.
[:SENSe]:FREQuency:CENTer:SET:STEP
Syntax
[:SENSe]:FREQuency:CENTer:SET:STEP
Function
Sets the current center frequency step to the center frequency.
Front Panel
FREQ, CF->Step
20.
[:SENSe]:FREQuency:CENTer:STEP:AUTO
Syntax
[:SENSe]:FREQuency:CENTer:STEP:AUTO OFF|ON|0|1
[:SENSe]:FREQuency:CENTer:STEP:AUTO?
Function
Enables or disables the Auto function of CF step.
The query returns “0” or “1”.
Default
ON|1
Front Panel
FREQ, CF Step, Auto/Manual
21.
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]
Содержание DSA1000A Series
Страница 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Страница 2: ......
Страница 14: ......
Страница 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Страница 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Страница 155: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 21 7 Execute results...
Страница 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...