RIGOL
Chapter 2 Command Systems
Programming Guide for DSA1000A series
2-12
markers in “Delta”, “Delta Pair” and “Span Pair” marker.
Explanations
<n> is is 1, 2, 3 or 4.
Sending :CALCulate:MARKer<n>:MODE selects a marker type
from “Delta”, “Delta Pair” or “Span Pair”.
The command is invalid in Zero-span mode.
Front Panel
Marker, Delta/Delta Pair/Span Pair
Marker –>, Mkr
Δ
->Span
17.
:CALCulate:MARKer:FCOunt:RESolution
Syntax
:CALCulate:MARKer:FCOunt:RESolution <freq>
:CALCulate:MARKer:FCOunt:RESolution?
Function
Sets the frequency counter resolution, the default unit is Hz, such as
:CALC:MARK:FCO:RES 1000 or :CALC:MARK:FCO:RES 1KHZ
The query returns “1000”
.
Explanations
<freq> ranges from 1 Hz to 100 kHz (1Hz, 10Hz, 100Hz, 1kHz,
10kHz, 100kHz)
Sending :CALCulate:MARKer:FCOunt:RESolution:AUTO can set
the resolution mode to Auto.
Default
1000 (1kHz)
Front Panel
Marker Fctn, Freq Count, Resolution, Manual
18.
:CALCulate:MARKer:FCOunt:RESolution:AUTO
Syntax
:CALCulate:MARKer:FCOunt:RESolution:AUTO OFF|ON|0|1
:CALCulate:MARKer:FCOunt:RESolution:AUTO?
Function
Set the resolution mode of frequency counter to Auto in order to get
a faster and accurately counting in connection with specified span.
The query returns “0” or “1”.
Default
ON|1
Front Panel
Marker Fctn, Freq Count, Resolution,
Auto
19.
:CALCulate:MARKer:FCOunt[:STATe]
Syntax
:CALCulate:MARKer:FCOunt[:STATe] OFF|ON|0|1
:CALCulate:MARKer:FCOunt[:STATe]?
Function
Turns on or off the frequency counter.
The query returns “0” or “1”.
Explanations
Sending :CALCulate:MARKer:FCOunt:X? queries the readout of
the frequency counter.
Содержание DSA1000A Series
Страница 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Страница 2: ......
Страница 14: ......
Страница 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Страница 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Страница 155: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 21 7 Execute results...
Страница 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...