Chapter 2 Command Systems
RIGOL
Programming Guide for DSA1000A series
2-99
1.
:SOURce:CORRection:OFFSet
Syntax
:SOURce:CORRection:OFFSet <rel_ampl>
:SOURce:CORRection:OFFSet?
Function
Sets the output amplitude offset for the tracking source in dB, such
as
:SOUR:CORR:OFFS 10
The query returns “1.01”.
Explanations <rel_ampl> ranges from -200 dB to 200 dB.
Default
0 dB
Front Panel
Source, TG Lvl Offset
2.
:SOURce:FREQuency:OFFSet
Syntax
:SOURce:FREQuency:OFFSet <freq>
:SOURce:FREQuency:OFFSet?
Function
Sets the output frequency offset for the tracking source, the default
unit is Hz, such as
:SOUR:FREQ:OFFS 1000000
The query returns “1000000”.
Explanations <freq> ranges from -300 MHz to 300 MHz.
Default
0Hz
Front Panel
Source, Freq Offset
3.
:SOURce:POWer:LEVel:IMMediate:AMPLitude
Syntax
:SOURce:POWer:LEVel:IMMediate:AMPLitude <ampl>
:SOURce:POWer:LEVel:IMMediate:AMPLitude?
Function
Sets the output power of the tracking source when the “Power
Sweep” is off, the default unit is dBm, such as
:SOUR:POW:LEV:IMM:AMP -10
The query returns “-1.01”.
Explanations <ampl> ranges from -20 dBm to 0 dBm when the parameter
<rel_ampl> in the :SOURce:CORRection:OFFSet command is 0, or
else the range should plus the value of <rel_ampl>.
Default
0 dBm
Front Panel
Source, Power Sweep, Off, TG Level
4.
:SOURce:POWer:MODE
Syntax
:SOURce:POWer:MODE FIXed|SWEep
Содержание DSA1000A Series
Страница 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Страница 2: ......
Страница 14: ......
Страница 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Страница 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Страница 155: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 21 7 Execute results...
Страница 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...