Communications
71
Approximate Values Between The DA Converter and Analog Output
Figure 6-4. Analog Output
6.11 Print Programming
Once a printer has been configured correctly in the setup mode, the instrument carries out the print functions; each
functioning mode has specific printouts, shown in the
. It is possible to define some
printing fields, the size of the characters, a heading and other options depending on the printer and the predefined
functioning mode. See
for connection information.
1. Turn on the instrument and press
as the firmware version displays.
displays.
2. Press
or
until
displays and then press
.
3. Press
or
until
displays and then press
.
4. Press
or
until
displays and then press
.
5. Press
or
until
displays and then press
.
6. Select the desired printer
and then press
.
7. Configure
,
,
,
and
(baud rate, number of transmission bits,
Set the analog output value when the scale displays zero weight (supplied when the scale is in underload);
This value can be between 0 and 65535 (values of the digital/analog converter); If a higher value is entered,
the instrument displays 0
Set the minimum value of the analog output:
With positive weight – the minimum value provided by the analog output, corresponding also to the under-
load condition.
With negative weight – the minimum value provided by the analog output, corresponding also to the over-
load condition.
This value can be between 0 and 65535 (values of the digital/analog converter); If a higher value is entered,
the instrument displays 0
DA Converter
Voltage
Current (mA)
1200
0
0
12700
--
4
58600
--
20
62650
10
--
Table 6-18. DA Converter and Analog Output
Parameter
Description
Table 6-17. Analog Output Parameters (Continued)
Ao = Analog output
Positive Weight
Negative Weight