3.2.8 Analog Output Menu
The ALGOUT menu is shown only if the analog output option is installed. If the analog output option is installed, configure all
other indicator functions and calibrate the indicator itself before configuring the analog output. See the
Analog Output
Card Installation Instructions
, PN 69089, for more information.
Figure 3-18. Analog Output Menu
ALG OUT Menu
Parameter Choices
Description
Level 2 submenus
SOURCE1
PROG
SCALE
n
Specifies the scale tracked by the analog output. PROG indicates that the analog output is under program control.
MODE1
GROSS
NET
Specifies the weight data, gross or net, tracked by the analog output.
OFFSET
0%
20%
Zero offset. Select 0% for 0–10 V or 0–20 mA output; select 20% for 4–20 mA output. This parameter must be set
before calibrating the analog output.
ERRACT
FULLSC
HOLD
ZEROSC
Error action. Specifies how the analog output responds to system error conditions.
Possible values are:
FULLSC: Set to full value (10 V or 20 mA)
HOLD: Hold current value
ZEROSC: Set to zero value (0 V or 4 mA)
MIN
000000
0–9999999
Specifies the minimum weight value tracked by the analog output. Specify a value in the range 0–9999999.
MINNEG
OFF
ON
Specify ON if the minimum weight (MIN parameter) is a negative value.
MAX
10000
0–9999999
Specifies the maximum weight value tracked by the analog output. Specify a value in the range 0–9999999
MAXNEG
OFF
ON
Specify ON if the maximum weight (MAX parameter) is a negative value.
TWZERO
40
0–65535
Tweak zero. Enter tweak value to adjust the analog output zero calibration. Use a multimeter to monitor the analog out
-
put value.
TWSPAN
59650
0-65535
Tweak span. Enter tweak value to adjust the analog output span calibration. Use a multimeter to monitor the analog
output value.
Table 3-11. Analog Output Menu Parameters
SCALES
SERIAL
FEATURE
PFORMT
SETPTS
DIG I/O
ALGOUT
VERS
number
40
number
SOURCE
PROG
OFFSET
20%
0%
ERRACT
SCALE n
FULLSC
HOLD
ZEROSC
TWZERO
59650
TWSPAN
MODE
GROSS
NET
If SOURCE1
≠
PROG
ALGOUT2
ALGOUT4
If dual-channel
card installed
number
10000
ON
MIN
number
000000
MINNEG
ON
OFF
MAX
OFF
MAXNEG