Configuration
© Rice Lake Weighing Systems
●
All Rights Reserved
23
DFSENS
2OUT
4OUT
8OUT
16OUT
32OUT
64OUT
128OUT
Digital filter cutout sensitivity. Specifies the number of consecutive readings that must fall outside the filter threshold
(DFTHRH parameter) before digital filtering is suspended.
for more information about digital filtering.
DFTHRH
NONE
2D
5D
10D
20D
50D
100D
200D
250D
Digital filter cutout threshold. Specifies the filter threshold, in display divisions. When a specified number of consecu
-
tive scale readings (DFSENS parameter) fall outside of this threshold, digital filtering is suspended. If NONE is
selected, the filter is always enabled.
for more information about digital filtering.
RATTRAP
OFF
ON
Enables RATTLETRAP
®
digital filtering. RATTLETRAP is most effective at filtering repeating vibrations caused by
mechanical noise from nearby machines but may increase settling times over standard digital filter selections.
SMPRAT
120HZ
240HZ
480HZ
960HZ
7.5HZ
15HZ
30HZ
60HZ
Sample rate. Selects measurement rate, in samples per second, of the analog-to-digital converter. Lower sample rate
values provide greater signal noise immunity: the default 120 Hz value may be too fast to provide the desired stability
in some static weighing applications.
NOTE: The maximum total sample rate for all configured A/D channels—the sum of the sample
rates for all scales—is 1200 Hz.
PWRUPMD
GO
DELAY
Power up mode. In GO mode, the indicator goes into operation immediately after a brief power up display test.
In DELAY mode, the indicator performs a power up display test, then enters a 30-second warm up period. If no motion
is detected during the warm up period, the indicator becomes operational when the warm up period ends; if motion is
detected, the delay timer is reset and the warm up period repeated.
TAREFN
BOTH
NOTARE
PBTARE
KEYED
Enables or disables push-button and keyed tares. Possible values are:
BOTH: Both push-button and keyed tares are enabled
NOTARE: No tare allowed (gross mode only)
PBTARE: Push-button tares enabled
KEYED: Keyed tare enabled
ACCUM
OFF
ON
Accumulator. Specifies whether the scale accumulator is enabled. If enabled, accumulation occurs whenever a print
operation is performed.
VISIBL
ON
OFF
Scale visibility. Specifies whether scale data is displayed.
PEAK HOLD
OFF
NORMAL
BI-DIR
AUTO
Peak hold. Used to determine, display, and print the greatest net weight read during a weighing cycle. The weighing
cycle ends when a print command is executed (AUTO setting) or when the peak weight is cleared by pressing
ZERO
or
. Press
GROSS/NET
to display gross weight data when using the peak hold function.
OFFPeak hold function is off.
NORMALPositive peak, manual reset. Greatest net weight is held in memory until the weight is removed from the
scale and either the
ZERO
or
key is pressed.
BI-DIRBi-directional peak, manual reset. Same as NORMAL, but peak value can be either positive or negative, deter
-
mined by absolute value.
AUTOPositive peak, auto print, auto reset. Automatic print occurs when scale load is 0 ± 10 display divisions and at
standstill. Following the print command, the peak value is cleared and reset automatically.
CALIBR
WZERO
WVAL
WSPAN
WLIN
REZERO
See Level 4 submenu descriptions in
.
SCALES Menu
Parameter
Choices Description
Table 3-2. SCALES Menu Parameters (Continued)