Chapter 5
5-24
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
REF 1029568 Rev B
Diagnostic Mode and Troubleshooting
68
UNABLE TO QUEUE DATA
FOR COMMUNICATIONS
Controller processor detected a RTOS queue full condition.
1. Replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
69
SAMPLING THREAD LOCKED
Controller processor detected that the sampling thread is locked.
1. Replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
70
EXECUTION THREAD LOCKED Controller processor detected that the main execution thread is locked.
1. Replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
71
INTERNAL RAM FAILURE
Controller processor detected an internal RAM memory checksum error.
1. Replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
72
UNABLE TO QUEUE DATA
FOR LOGGING
Controller processor detected a RTOS queue full condition.
1. Replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
73
Unused
74
Unused
75
TCB IPC MESSAGE TOO BIG
Controller processor detected an IPC message that was too big.
1. Replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
76
TCB BUZZER SUPERCAP
FAILURE
Controller processor detected that the backup alarm buzzer capacitor test failed
a limits check for greater than five minutes.
1. Enter Diagnostics mode and use the System Test screen to check if the
backup alarm can be turned on and off.
2. If the problem persists replace the controller PCB.
3. Retrieve the DRPT and contact technical support.
77
HIGH PRESSURE CONDITION
Controller processor detected a high pressure condition and annuciated the
Hi P
Reg
alarm. No action necessary. If this condition persists for greater than 10
seconds the system declares error code 41
78
LOW PRESSURE CONDITION
Controller processor detected a low pressure condition and annuciated the
Lo P
Reg
alarm. No action necessary.
79
UNABLE TO MAINTAIN
PRESSURE SUPPORT
Unable to maintain Pressure Support. System error declared after condition is
present for one minute.
1. Retrieve the DRPT and contact technical support.
80
UNABLE TO QUEUE DATA TO
IPC
Controller processor detected a RTOS queue full condition.
1. Replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
81
INVALID IPC MESSAGE ID
Controller processor detected an IPC message with an invalid message id.
1. Replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
Table 5-3: Significant Event Codes (Continued)
Event ID
Comment / Corrective Action