Chapter 7
Testing, Calibration, and Performance Verification
REF 1029568 Rev B
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
7-15
8. Connect the pressure tubing from the oxygen port of the oxygen
enrichment adapter to the pneumatic calibration analyzer
(Figure 7-12).
Figure 7-12: Pressure Calibration Setup
9. On the service PC, open the Service Calibration Software and start the
calibration by clicking on the white Run arrow in the upper left corner
of the window (Figure 7-13).
Figure 7-13: Starting Calibration
Run
arrow