Chapter 7
7-10
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
REF 1029568 Rev B
Testing, Calibration, and Performance Verification
7. When the Success window appears, click
OK
Figure 7-7: Calibration Success Window
7.5.3
If Real-Time Clock Calibration Fails
If calibration fails, confirm that ventilator time and date are set to within 5
minutes of the service PC. Re-run the calibration. If the calibration fails again,
replace the controller PCB and retest.