Chapter 8
8-6
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
REF 1029568 Rev B
Component Removal/Installation
8.3 Display Assembly
Follow these steps to remove or replace the display assembly (Figure 8-6,
Figure 8-7, Figure 8-8, Figure 8-9, Figure 8-9). Reverse to install.
1. Remove top cover (section 8.2).
2. Disconnect the VGA/UI ribbon cable from the user interface PCB.
3. Disconnect the ground cable from the user interface PCB.
Figure 8-6: Display Assembly and User Interface PCB
UI/VGA ribbon cable
UI/VGA ground cable
Display assembly
User interface PCB