Chapter 7
Testing, Calibration, and Performance Verification
REF 1029568 Rev B
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
7-19
21. Click
OK
to begin the flow calibration.
22. Note the flowmeter reading: click
INCREASE BLOWER SPEED
or
DECREASE BLOWER SPEED
to achieve a flow of 150 to 160 LPM,
then click
DONE
Figure 7-19: Increase/Decrease Flow
23. Note the flowmeter reading and adjust the flow control valve to
achieve a flow of 146 LPM, then click
OK
.
24. The screen then prompts you to adjust the flow valve to achieve
various flow settings. Ensure that each flow setting is achieved before
clicking
OK
.