Intro
Read This
First
Transmitter
Installation
Power Supply
Wiring
I/O
Wiring
Batch
Wiring
Startup
Trouble-
shooting
Program-
ming
Main
Menu
Config
Menu
System
Menu
Model
Numbers
Xmitter
Specs
Return
of Goods
Calibration
Menu
Diagnostic
Menu
Datamate 2200 O&M
Page 61
Menu or Display Description
Electronic Zero
Displays the current electronic zero and the standard deviation for the last
1. Elec. Zero
automatic zero. With the correct access (level 2) the user can modify the
1.
Params
displayed ZERO value with the ^ and > keys.
Electronics Zero
[ SX.XXXXeSXX]
DEG
Pressing ENTER will display the standard deviation (StdDev) of the last
automatic zero. The StdDev cannot be modified by the user.
StdDev
[ SX.XXXXeSXX]
DEG
Electronic Zero
Displays the current zero and the standard deviation for the zero.
2. Zero Electronics ZERO: SX.XXXXeSXX DEG
SDEV: X.XXXXeSXX DEG
ENTER to Calibrate
CANCEL to Exit
Choices from this menu are ENTER to re-zero and CANCEL to exit.
While the Zero is performed the display will read:
Zeroing
When Zeroing is complete the new zero and the standard deviation is
displayed:
ZERO: SX.XXXXeSXX DEG
SDEV: X.XXXXeSXX DEG
CANCEL to Exit
Press CANCEL to Exit and remove any jumper at the input of the Datamate
2200
Zero
Displays the current transducer zero. The transducer zero is the zero offset
2. Transducer Zero contribution of the transducer and installation. The transducer zero is
determined by subtracting the electronic zero from the system zero.
Press ENTER to review and/or modify the transducer zero.
Transducer Zero
With the correct access (level 2) the user can modify the displayed ZERO
value with the ^ and > keys.
Transducer Zero
[SX.XXXXeSXX]
DEG
Zero
This menu allows the user to review and modify the System zero of the
3. System Zero
Datamate 2200.
The System zero is the combined zero offset of the electronics, transducer
and the installation when there is no flow.
Choices from this menu are Syst Zero Params, Zero System, and Exit.
System Zero
Displays the current system zero and the standard deviation for the last
1. Syst Zero
automatic zero. Neither value can be modified from this menu
1.
Params
System Zero
[ SX.XXXXeSXX]
DEG
Pressing ENTER will display the standard deviation (StdDev) of the last
automatic zero. The StdDev cannot be modified by the user.
StdDev
[ SX.XXXXeSXX]
DEG