2.1 Transmitter Area Classification
WARNING: IMPROPER INSTALLATION IN A HAZARDOUS ENVIRONMENT COULD
CAUSE AN EXPLOSION. INSTALL THE Datamate 2200 IN AN AREA COMPATIBLE
WITH THE HAZARDOUS ENVIRONMENT SPECIFIED ON THE APPROVALS TAG.
The Datamate 2200 has been tested by an independent third party laboratory and has
been found to meet the following standards:
FCC Part 15, Class A - Emissions
EN61326 (1977), Class A - Emissions
EN61326 - Industrial Location Immunity
Hazardous Area Approval:
CSA
Class I, Div 2, Group C and D
Class II, Div 2, Groups E, F, and G
Class III
2.2 Orientation and Mounting
Orient the transmitter so that the enclosure cover and the conduit openings are easily
accessible.
2.3 Temperature, EMI/RFI, Humidity, Vibration and Mechanical Shocks
The specified limits for the transmitter are as follows:
• Ambient temperature (with optional LCD) -4 to 140°F (-20 to 60°C)
• Ambient temperature (without optional LCD) -40 to 140°F (-40 to 60°C)
• EMI/RFI Class A digital device, pursuant to Part 15 of the FCC Rules
• Humidity: 10 to 100% condensing
• Vibration [Nonoperating 5g, 3 Axis, 10 sweeps from 10-800 Hz]
[Operating 2.5g, 3 Axis, 10 sweeps from 10-800 Hz]
• Mechanical Shocks [Nonoperating 100 shocks up to 30g/11ms]
[Operating 100 shocks up to 15g/11ms]