UM-1096
Revision A
19
Chapter 6
Maintenance Procedure
6.1
Regular Maintenance
RC Systems recommends performing calibrations at regular intervals to ensure proper functionality of
the SensePoint. During routine calibration, RC Systems recommends a visual inspection of sensor head,
enclosure and conduit entries to check for cleanliness and physical integrity. Cleaning the detector is
recommended when necessary but be aware that some cleaning compounds may be detected by an
operational detector depending on the sensor type. So, proper precautions should be taken.
RC Systems recommends calibrations:
Immediately prior to placing a gas detector in service
Any time a new sensor is installed
Every six months for routine calibrations (more often if sensor is known to have been
exposed to gas for extended periods of time)
Periodic bump tests are recommended if detector has potentially been exposed to
incompatible gases to ensure correct operation
6.2
Sensor Replacement
When a sensor has reached its end of life, it is necessary to replace the sensor. For sensor replacement
Содержание 10-0517
Страница 4: ...UM 1096 Revision A III Drawings 44...
Страница 10: ...UM 1096 Revision A 5 Figure 2 Remote Sensor Enclosure Mounting Dimensions...
Страница 26: ......
Страница 49: ...UM 1096 Revision A 44 Appendix 9 Drawings Drawings Figure 13 Mounting Dimensions...
Страница 50: ...UM 1096 Revision A 45 Figure 14 10 0517 CPU Board Figure 15 10 0533 I O Board...
Страница 51: ...UM 1096 Revision A 46 Figure 16 10 0532 Modbus Relay Board...