Instructions for Use MPR1 DATALOGGER
Manufacturer:
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
Page 91 of 98
Guidelines and manufacturer declaration - electromagnetic immunity
The MPR1
is intended for use in an electromagnetic environment as specified below. The customer or
user of the MPR1 should ensure that it is used in such an environment.
Conducted RF
disturbance
variable as per
IEC 61000-4-6
Radiated RF
disturbance
variable as per
IEC 61000-4-3
3 V
eff
150 kHz to
80 MHz
3 V/m 80 MHz to
2.5 GHz
3 V
eff
150 kHz to
80 MHz
3 V/m 80 MHz to
2.5 GHz
Portable and mobile RF communication
equipment should not be used closer to
any part of the device (including cabling
and sensors) than the recommended safe
distance, which is calculated using the
relevant equation for the transmission
frequency.
d = 1.17 * P
1/2
d = 1.17 * P
1/2
for 80 MHz to 800 MHz
d = 2.33 * P
1/2
for 800 MHz to 2.5 GHz
d: distance in m
P : transmitted power in W
The field strength of stationary radio
transmitters should be less than the
conformity level at all frequencies, as
demonstrated by an on-site
a
investigation.
b
Interference is possible in the area
surrounding devices carrying the following
symbol.
At 80 MHz and 800 MHz the higher frequency range applies.
These guidelines may not be applicable in all cases. The propagation of electromagnetic waves is
influenced by their absorption and reflection by buildings, objects and people.
a
The field strength of stationary emitters, such as mobile phone network base stations and mobile
agricultural broadcasting equipment, amateur radio stations, AM and FM radio and television
broadcasters cannot be accurately defined in advance. To determine the electromagnetic environment in
respect of any emitters, a study of the site should be consulted. If the measured field strength at the site
in which the MPR1 is used, exceeds the above conformity threshold, the MPR1 should be monitored, to
demonstrate that it is functioning correctly. If unusual characteristics are observed, additional measures
may be required, e.g. a changed alignment or another site of use of the MPR1.
b
The field strength should be less than 3 V/m in the frequency range 150 kHz to 80 MHz.