Telemetry channel pressure
Number of channels:
1
Measuring site designations:
ICPT
Measurement ranges;
- 40 ... + 400 mm Hg
Pressure value resolution:
1 mm Hg
Accuracy:
± 2 mm Hg
Sampling rate:
5 Hz
Defibrillator protection
yes
Analogue outputs
Number of channels:
2
Transferable input values:
Invasive pressure twice
Characteristic features:
5
µ
V / V / mmHg
Measuring range:
- 40 ... + 360 mmHg
Resolution:
< 0.5 mmHg, typical 0.23 mmHg
Accuracy after zeroing*1:
± 1.5 mmHg in tempered condition (25
°
Celsius)
Output frequency:
100 Hz
Thermal drift:
maximum 0.5 mmHg / K
*1: Zeroing of the downstream monitor
Interface for the expansion of the apparatus
Type:
RS232 (V24) + power supply
Application:
unit service and optional extension of the apparatus
PC interface
Type:
USB
Application:
connection of laptop
General data
Type designation:
RAUMEDIC MPR 1 DATALOGGER
Mode of operation:
continuous operation
Volume of alarm:
minimum 45 dB(A) typical 60 dB(A), 1 m, in stand holder
Display:
Black/white LCD 320 x 240 pixel (1/4 VGA), 74 x 97 mm
Trend display:
optional 4 min., 8 min., 20 min., 40 min., 80 min. or 4 h
Alarm limits:
to be selected by menu
Language:
German, other language versions optional
Dimensions:
approx. 200 x 69 x 150 mm (W x D x H)
Mass:
approx. 0.95 kg with battery
Fuse in battery compartment *
2
:
type: TR5 / No. 372 / 250V / 1.25 AT / Wickmann
Power supply:
internal battery
Type of battery:
NiMH, Panasonic HHR210AH-1Z
Operating period:
> 12 h (charged, new battery pack, all channels occupied,
illumination off)
Charger / mains unit:
Wide range adapter Art. No. 284007-002 or Art. No. 284027-001
Input:
100-240 V / 50-60 Hz / 400-150 mA,
Output: 12 V / 1.5 A, protection class II, IP 40
Class of protection:
- class of protection II with internal and external power supply
- type CF IEC 60601-1
- Directive 93/42/EEC Class II b
IP protection class:
IP 21 in operation (sockets to the bottom, maximum inclination
20°)
*2: The fuse may be replaced by trained staff only. If the apparatus does not operate after the first replacement, remove the fuse again
and take the apparatus to the after-sales service department for repair. The fuse can be obtained from the manufacturer, if
necessary.