Instructions for Use MPR1 DATALOGGER
Manufacturer:
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
Page 7 of 98
•
In all fastening and erection variants, please make sure that the apparatus cannot drop or cause other
hazards. Please also ensure a safe laying of the cables so that a tipping over of the arrangement is
prevented.
•
Do not use the apparatus in potentially explosive surroundings. The apparatus shall have to be set up
at a minimum distance of 25 cm to any anaesthesia units and gas filled hoses.
•
After use the accessories shall have to be cleaned, disinfected or sterilised in keeping with the
instructions for use. If the accessory has been provided with separate instructions for use, the
instructions rendered there shall be applicable.
•
If the apparatus has been taken from colder into warmer surroundings, it may be used only when the
temperature of the apparatus and of the accessories has adapted to the room temperature; there is no
danger of condensation water forming or any condensation water already developed has completely
dried.
•
Please also observe the safety instructions in conjunction with the use of sensors and transducers
which have been included to the respective sections of the instructions for use.
•
Note that the optional accessory reader, TDT1 readP, can only be used for telemetric pressure
measurement with a pressure measurement implant.
•
No liquids or fluids shall enter the apparatus. If this should happen nonetheless, first remove the mains
adapter from the socket. Hand over the apparatus to the after-sales service department for inspection.
Subsequently a safety-related check is necessary.
•
The MPR1 has been designed in conformity with standard IEC 60601-1. It is a product of class of
protection II with an internal power source and an external mains adapter, and has been allocated to
class IIb.
•
If the apparatus is subjected to strong electro-magnetic fields, please observe that the pressure display
may vary by up to 2 mmHg. For this reason, avoid using mobile phones, therapeutical microwave
devices or similar units in the vicinity of the apparatus.
•
The following instructions shall have to be observed for the installation of the system (with laptop and/or
foreign unit – monitor):
•
Moveable multiple sockets must not lay on the floor.
•
Additional moveable multiple sockets or extension cables must not be connected up to the system.
•
Apparatuses, which are not part of the system, must not be connected.
•
The moveable multiple sockets have to be suitable for the load of the system.
•
The moveable multiple sockets which are supplied along with the system only may be used for the
supply of the devices of the system.
•
Observe the instructions for the installer on how to set up the system in an ideal way. Please also
read chapters 3.5, 3.6 and 3.7.