Instructions for Use MPR1 DATALOGGER
Manufacturer:
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
Page 56 of 98
4.4.5.5 Measuring the CPP using the ICPT
Determination of the CPP takes place as described under point
Measuring the CPP. In this
respect, the individual ICP pressure sensors have the following fixed priorities for determination of the
CPP:
•
ICP1 - highest priority
•
ICPT - lowest priority
Cabled pressure measurement of the ICP has priority over telemetric pressure measurement of the ICP.
•
The CPP is determined from ART and ICP1, if ICP1 is available.
•
The CPP is determined from ART and ICP2, if ICP2 is available and ICP1 is not.
•
The CPP is determined from ART and ICPT, if neither ICP1 or ICP2 are available and ICPT is available.
4.5
Configuring the graphics display
Screen
as well as screen renders graphical displays of the measured or calculated parameters;
screen renders live data, and screen provides the trend. If several transducers have been connected,
the display area is divided up into several tracks. Thus, an overview of all connected parameters is obtained.
At the same time, however, the amplitude resolution decreases with the increase in the number of tracks.
Sometimes, however, it is necessary to have certain or individual parameters in a higher resolution
amplitude. In screen as well as in screen it is possible to deselect tracks in the graphics display after
having pressed the Shift key of the connected (e.g. ICP, ART) or calculated signals (CPP), which increases
the resolution of the remaining tracks; in addition, tracks can be added, which decreases the resolution of
the tracks.
The key function is alternating for the deselection of the tracks. Available parameters for the function
keys are displayed. The graphics track can be deselected by pressing the key, which will be added
again when the key is pressed again.
From the following example, you can see which operating steps are required to configure the graphical
representation.
Step 1: Press the Shift key, and
thus change the allocation of the F
keys; the parameters available for
the F keys are displayed.
Step 2: Select the F key
parameters which you want to
select or deselect as graphics
tracks. For example, press the F
keys ART, ICP1 and ICP2 to
deselect these graphics tracks and
to display the CPP at maximum
resolution.
Step 3: Either you press the Shift
key again to return to scroll mode
(as in step 1) or it is set
automatically again after a few
seconds.
Tracks can be deselected until no curve is visible any more. In this case, the display will render “With
Shift + channel name you can switch between curves”. The procedure in screen 2 and 3 is identical.
This
document
is
for
information
only