Remote Control Commands
R&S
®
NGL200/NGM200
96
User Manual 1178.8736.02 ─ 06
SINK
If respective channel operation mode is detected in sink mode,
corresponding trigger-out parameters are triggered.
PROTection
If respective channel operation mode is detected in protection
mode (OVP, OCP, OPP OTP), corresponding trigger-out param-
eters are triggered.
Example:
TRIG:SOUR OMOD
TRIG:SOUR:OMOD:CHAN OUT1
TRIG:SOUR:OMOD CV
If CV mode is detected on Ch1, a trigger event is activated to
corresponding trigger-out parameters.
TRIGger[:SEQuence][:IMMediate]:SOURce:OMODe:CHANnel
<arg0>
TRIGger[:SEQuence][:IMMediate]:SOURce:OMODe:CHANnel?
<arg0>
Sets or queries the device channel for trigger source "Operation Modes".
See
Parameters for setting and query:
<arg0>
OUT1 | OUTP1 | OUTPut1 | OUT2 | OUTP2 | OUTPut2
OUT1 | OUTP1 | OUTPut1
Ch1 is selected as the device channel for trigger source.
OUT2 | OUTP2 | OUTPut2
Ch2 is selected as the device channel for trigger source.
Example:
TRIG:SOUR:OMOD:CHAN 1
Ch1 is selected as the device channel for trigger source "Opera-
tion Modes".
TRIGger[:SEQuence][:IMMediate]:SOURce:OUTPut:CHANnel
<arg0>
TRIGger[:SEQuence][:IMMediate]:SOURce:OUTPut:CHANnel?
<arg0>
Sets or queries the device channel for trigger source "Output".
See
Parameters for setting and query:
<arg0>
OUT1 | OUTP1 | OUTPut1 | OUT2 | OUTP2 | OUTPut2
OUT1 | OUTP1 | OUTPut1
Ch1 is selected as the device channel for trigger source.
OUT2 | OUTP2 | OUTPut2
Ch2 is selected as the device channel for trigger source.
Example:
TRIG:SOUR:OUTP:CHAN 1
Ch1 is selected as the device channel for trigger source "Out-
put".
Trigger Commands