Instrument Functions
R&S
®
NGL200/NGM200
41
User Manual 1178.8736.02 ─ 06
The R&S
NGL/NGM outputs the set voltage of the selected channel and displays
the corresponding values in the home window.
For more information on the operation modes, see
Figure 6-2: Voltage and current settings in the instrument
6.2
Activating the Channels Output
The outputs of all the channels (Ch1, Ch2) can be switched on or off by toggling the
[Output] key on the front panel.
By default, the output is turned off when the instrument is switched on. The output is
also automatically turned off when no channel is selected. This design prevents a con-
nected load from being damaged unintentionally.
1. Press the required channel key.
Selected channel key (Ch1, Ch2) illuminates.
2. Press [Output] key.
The R&S
NGL/NGM outputs the set voltage of the selected channel.
Depending on the operating mode, the font text in the channel display area shows
green in CV mode, red in CC mode and blue in CR mode.
Note: The R&S
NGL/NGM goes into CR mode when CR is enabled and the instru-
ment operates in sink mode.
See
Chapter 5.5, "Operation Modes"
Multiple outputs can be turned on or off at the same time.
Activating the Channels Output