Remote Control Commands
R&S
®
NGL200/NGM200
106
User Manual 1178.8736.02 ─ 06
// by the set 1 Ampere
CURR:STEP?
// queries the current step size
// response: 1.000
[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement]
............................................. 106
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
....................................................... 106
[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement]
<stepsize>
[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement]?
[<stepsize>]
Sets or queries the incremental step size for the
Setting parameters:
<stepsize>
numeric | DEFault
numeric
Step value in A.
DEFault
Default value of stepsize.
Range:
0.0001 to 2.000
Increment: 0.0001
*RST:
0.010
Default unit: A
Query parameters:
<stepsize>
DEFault
Queries the default voltage step size.
Example:
INST OUT1
CURR:STEP 0.005
CURR:STEP DEF
VOLT:STEP? DEF
-> 0.1000E+00
Returns the default stepsize for current.
See
Example "Configuring the current output"
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
<current>
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?
[<current>]
Sets or queries the current value of the selected channel.
Setting parameters:
<current>
numeric | MIN | MAX | UP | DOWN
numeric
Numeric value in the range of 0.000 to 6.010.
MIN
Minimum current at 0.010 A.
Configuration Commands