Instrument Functions
R&S
®
NGL200/NGM200
71
User Manual 1178.8736.02 ─ 06
4. Select "Enabled" to activate the "FastLog" function.
The R&S NGM starts the fast logging and displays the "Fast Logging" icon at the
device status bar information.
5. Select "Triggered" to "ON" if you want "FastLog" to be enabled by a trigger event.
6. Select "Write to USB" if you want to save the measurements data to a binary file.
The file is saved to the directory specified in the "Target Folder". The R&S NGM
creates a file for each channel and logs the measurement data to the binary file.
7. Select "Write to SCPI" to transfer the data to a SCPI client.
Note: The sum of all sample rates have to be smaller or equal to 500kS/s, i.e. sam-
ple rate of Ch 1 + sample rate of Ch 2 <= 500kS/s.
6.12
CSV Settings
A CSV file stores tabular data (numbers and text) in plain text. Each line of the file is a
data record and each record consists of one or more fields, separated by a file delim-
iter. The "CSV Settings" provides you ways to format the fields that are stored in the
data logging. See
CSV Settings