Instrument Functions
R&S
®
NGL200/NGM200
44
User Manual 1178.8736.02 ─ 06
3. Activate the "Fast Transient Response" from the menu.
The R&S
NGL/NGM applies the fast transient response on the operating condition
and displays the "Fast Transient Response" icon on the selected channel status
bar information.
The setting of "Fast Transient Response" affects the output impedance control loop
behavior. See
6.2.3
Output
The "Output" menu provides the settings for
,
and
on the output mode.
1. Press [Settings] key.
The R&S
NGL/NGM displays the device/channel menu window.
2. Select the required channel tab to apply output delay.
The R&S
NGL/NGM displays the selected channel menu.
3. Select the "Output" menu item.
The R&S
NGL/NGM displays the "Output" dialog.
Figure 6-6: Output delay dialog
6.2.3.1
Impedance
The output impedance function is disabled during voltage ramp time of "Ramp" func-
tion. See
1. Select the "Impedance" menu item to configure the required values.
The R&S
NGL/NGM displays the "Output Impedance" dialog.
Activating the Channels Output