Remote Control Commands
R&S
®
NGL200/NGM200
120
User Manual 1178.8736.02 ─ 06
Usage:
Event
[SOURce:]CURRent:PROTection:DELay:INITial
<duration>
Sets the initial fuse delay time once output turns on.
Parameters:
<duration>
numeric | MIN | MAX
numeric
Numeric value for initial fuse delay.
MIN
Min value for initial fuse delay.
MAX
Max value for initial fuse delay.
Range:
0.00 to 60.00
*RST:
0
Default unit: s
Example:
See
[SOURce:]CURRent:PROTection:DELay
<duration>
Sets the fuse delay time.
Parameters:
<duration>
numeric | MIN | MAX
numeric
Numeric value for the initial fuse delay.
MIN
Min value for initial fuse delay.
MAX
Max value for initial fuse delay.
Range:
0.00 to 10.00
*RST:
0
Default unit: s
Example:
See
[SOURce:]CURRent:PROTection:LINK
<arg0>
[SOURce:]CURRent:PROTection:LINK?
<arg0>
Sets or queries the fuses of several selected channels (fuse linking).
Parameters for setting and query:
<arg0>
1 | 2
Configuration Commands