Remote Control Commands
R&S
®
NGL200/NGM200
150
User Manual 1178.8736.02 ─ 06
Example:
LOG:COUN MAX
LOG:COUN? MAX
-> 10000000
LOG:DURation
<span>
LOG:DURation?
[<span>]
Sets or queries the duration of the data logging.
Setting parameters:
<span>
numeric_values | MIN | MAX
numeric_values
Duration of the data logging captured in the range of 0 s to
3.49*10^5 s.
MIN
Minimum duration of the data logging captured at 0 s.
MAX
Maximum duration of the data logging captured at 3.49*10^5 s.
Default unit: s
Query parameters:
<span>
MIN | MAX
Returns the duration of the data logging.
Example:
LOG:DUR MAX
LOG:DUR? MAX
-> 349000
LOG:FNAMe?
Queries the filename and storage location for the data logging.
Example:
LOG 0
LOG:FNAM?
-> ""
LOG 1
LOG:FNAM?
-> "/int/logging/log-20190318T1141853.407.csv"
Enables the data logging and queries the data log filename.
Usage:
Query only
LOG:INTerval
<interval>
LOG:INTerval?
[<interval>]
Sets or queries the data logging measurement interval. The measurement interval
describes the time between the recorded measurements.
Setting parameters:
<interval>
numeric_value | MIN | MAX
numeric_value
Measurement interval in the range of 0.1 s to 600 s.
Data and File Management Commands