Remote Control Commands
R&S
®
NGL200/NGM200
133
User Manual 1178.8736.02 ─ 06
ARBitrary:DATA
<data>
Sets or queries the arbitrary points for the previous selected channel. Max. 4096 arbi-
trary points can be defined. The dwell time between 2 arbitrary points is specified from
1 ms to 20 days.
Parameters:
<data>
voltage1, current1, time1, interpolation mode1, voltage2, cur-
rent2, time2, interpolation mode2, ...
Voltage and current settings depending on the instrument type.
If the interpolation mode is sets to 1, it indicates that the mode is
activated. If the interpolation mode is sets to 0, it indicates that
the mode is not activated.
Example:
INST OUT1
ARB:DATA 10,1,0.5,0
Defines one arbitrary point with: Voltage1 = 10 V and Current1 =
1 A, Time1 = 500 ms and Interpolation mode1 = 0 (disabled).
ARB:DATA?
-> 10.000, 1.000, 0.50, 1
Returns defined arbitrary points for the previous selected chan-
nel.
See
Example "Configuring an arbitrary sequence"
ARBitrary:FNAMe
<filename>[,<location>]
ARBitrary:FNAMe?
[<location>]
Sets or queries the file name and storage location for the arbitrary function.
Parameters for setting and query:
<filename>
Filename of the arbitrary function.
<location>
INT | EXT | DEF
INT
Internal memory
EXT
USB stick
DEF
Internal memory
Example:
ARB:FNAM "01.CSV"
ARB:FNAM? INT
-> "01.CSV"
ARBitrary:LOAD
Loads an arbitrary table from a file (filename specified with
ARB:FNAM
)
Advanced Operating Commands