Instrument Functions
R&S
®
NGL200/NGM200
53
User Manual 1178.8736.02 ─ 06
Figure 6-13: Battery model editor dialog
1. Configure the "Battery Model Editor" with the required state of charge (SoC), open-
circuit voltage (Voc) and internal resistance (ESR).
The R&S NGM displays the on-screen keypad for data entry.
See also
2. Confirm values with the unit keys.
3. Select the "Plus" or "Minus" icon to add or delete the battery model data from the
dialog.
● "Plus": A new row of battery model data is added to the end of the table. The
new addition is a copy of the last battery model data in the table.
● "Minus": To delete, select the row of battery model data for deletion follows by
the "Minus" icon.
4. Set the "Capacity" and "Initial SoC" to configure the battery capacity and initial
state of charge settings.
The R&S NGM displays the on-screen keypad for data entry.
5. Select (New file) or (Existing file) to save the battery model data.
The R&S NGM displays a popup message to shows that file saved successfully.
Battery Simulator