MANUALE ISTRUZIONI / INSTRUCTION MANUAL
Modello/Model R221/A - R221/A DHHS
C
ALIBRATION
21
R22
1/A
- R221
/A DHHS -
C
alib
ration
Ralco
srl
• Remove the part of the cover necessary to access the screws, see Chapter -
C
OVER
R
EMOVAL
• Remove the light source protection heatsink by unscrewing the fixing screws . This allows
you to access the light source.
• If the light-field needs to be moved laterally, loosen (not remove) the fixing screws
A
.
• Adjust through screw
B.
• When calibration is terminated, lock the screws
A.
See the following
Fig. Light Field Adjustment
Vertical Alignment
• If the light field is
smaller
than the X-ray field, lower the lamp by adjusting screw
C.
• If the light field is
bigger
than the X-ray field, raise the lamp by adjusting screw
C
.
See
Fig. Light Field Adjustment.
Fig. Light Field Adjustment)
Transversal Calibration (CROSS)
If the light-field needs calibration, the mirror
needs to be adjusted as follows:
• Remove the part of the cover
necessary to access the screws, see
Chapter -
C
OVER
R
EMOVAL
.
• Loosen the mirror fixing screw
A
(not
remove) and shift it to adjust the
position of the mirror, see
Fig.
Transversal Calibration (CROSS).
• Once you have regulated the mirror
tighten the screw
A
and remount the
cover, see Chapter -
C
OVER
R
EMOVAL
.
Fig. Transversal Calibration (CROSS)