MTR221/A - MTR221/A DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
Modello/Model R221/A - R221/A DHHS
O
PERATION
I
NSTRUCTIONS
17
R22
1/A
- R221
/A DHHS -
Ope
ra
tion Instru
ctio
ns
Ralco
srl
OPERATION INSTRUCTIONS
L
IGHT
/X-
RAY
F
IELD
S
ETTING
• Set the collimator at 100 cm SID.
• By rotating the knobs and with the knob index on the front panel, follow the tags of the
scale.
• Do not force the knobs.
• Activate the light field by pressing the related push-button on the collimator front panel.
• The collimator is ready to operate.
The collimator has been designed to operate as follows:
• Supply constantly connected during operation of the equipment.
• Light source ON time for the light field is factory set at 30 seconds (tol.20%). It is adjustable
from 30 to 45 seconds via a trimmer on the timer board.
• The normal ON/OFF cycle is set at 1 minute followed by an idle time of 4 minutes to allow
for cooling time (i.e. 1 minute ON/ 4 minutes OFF).
• The field is set by the two knobs on the front panel by following the indications of the scale
with the knob index or visually with the light field.
Front Panel
WARNING
:
P
ROLONGED
LIGHTING
WITHOUT
ALLOWING
LIGHT
SOURCE
TO
COOL
CAUSES
THE
COLLIMATOR
TO
OVERHEAT
IN
THE
AREA
NEAR
THE
LIGHT
SOURCE
.
F
OR
EMERGENCIES
:
MAXIMUM
LIMIT
ADIVSED
IS
5
SUCCESSIVE
LIGHT
SOURCE
OPER
-
ATIONS
. A
LLOW
THE
COLLIMATOR
TO
COOL
(
ABOUT
10
MIN
).
B
Y
INCREASING
THE
CONSECUTIVE
"ON"
TIMES
,
THE
SUBSEQUENT
REST
PERIODS
NEED
TO
BE
EXTENDED
ACCORDINGLY
. I
T
IS
IMPRUDENT
TO
PERFORM
TWO
CONSEC
-
UTIVE
CYCLES
OF
5
ON
TIMES
.