MTR221/A - MTR221/A DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
Modello/Model R221/A - R221/A DHHS
O
PTIONAL
I
TEMS
38
R22
1/A
- R221
/A DHHS -
Option
al Ite
m
s
Ralco
srl
OPTIONAL ITEMS
This collimator may have the following optional items; a detailed description is provided in this
chapter.
RO 001/A Fixed Mounting Flange
Thickness of this mounting flange is 20 mm.
When the collimator is assembled check the collimator to Focal Spot Alignment
(Primary Shutter Cut-Off).
Inspect the four images of the four collimator shutters which form the edges of the x-ray field.
A definitely indistinct edge indicates that the primary shutter, close to the focal spot, is the one
forming the line, rather than the outermost shutter.
To correct the condition, use the four mounting/centering adjustment screws to shift the
collimator in the direction of the indistinct line. Repeat the test film exposure after making the
adjustment.
T
HE
HEEL
EFFECT
WILL
CAUSE
THE
FIELD
TOWARD
THE
CATHODE
TO
BE
SLIGHTLY
LESS
SHARP
THAN
ON
THE
OTHER
THREE
SIDES
.
THIS
IS
NORMAL
AND
CANNOT
BE
CORRECTED
BY
ADJUSTMENT
.
IN
ADDITION
,
AN
X
-
RAY
TUBE
OF
12°
OR
LESS
TARGET
ANGLE
WILL
PRODUCE
AN
ASYMMETRICALLY
SHAPED
FIELD
WHEN
A
LARGE
FIELD
SIZE
IS
USED
AT
SHORT
SID
,
BECAUSE
OF
ANODE
CUT
-
OFF
EFFECT
.
THIS
IS
NORMAL
AND
MAY
NOT
BE
CORRECTED
BY
ADJUSTMENT
.