Vertical Alignment
• Remove the part of the cover necessary to access the screws, see Chapter -
C
OVER
R
EMOVAL
.
• If adjustment is required loosen the two screws
C
holding the light support.
• If the light-field is smaller than the X-ray field, lower the light source by adjusting screw
D
.
• If the light-field is bigger than the X-ray field, raise the light source by adjusting screws
D
.
• Tighten the two screws
C
. See
Fig. Light Field Adjustment
Longitudinal Alignment (Long)
• Remove part of the cover, see Chapter -
C
OVER
R
EMOVAL
.
• If the light-field needs to be moved laterally, loosen (not remove) the fixing screws
A.
• Use screw
B
to adjust transversely.
• When calibration is terminated, lock the screws
A.
See
Fig. Light Field Adjustment
Fig. Light Field Adjustment