
MTR221/A - MTR221/A DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
Modello/Model R221/A - R221/A DHHS
C
OMPLIANCE
V
ERIFICATION
26
R22
1/A
- R221
/A DHHS -
C
omplia
nce V
erification
Ralco
srl
Electromagnetic Immunity for All Equipment and Systems
Collimator R221/A - R221/A DHHS is intended for use in the electromagnetic environmenmt specified below. The customer or the
user should assure that it is used in such an environment.
Immunity
Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
EN 60601-1-2
test level
Hospital
Electronical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for input/output lines
>3m
EN 60601-1-2
test level
Hospital
Surge
IEC 61000-4-5
1 kV differential mode
2 kV common mode
EN 60601-1-2
test level
Hospital
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
0% U
n
for 0.5 cycles.
40% U
n
for 5 cycles
70% U
n
for 25 cycles
0% U
n
for 5 s
EN 60601-1-2
test level
Hospital
Power frequency (50/
60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
EN 60601-1-2
test level
Hospital
Содержание MTR 221/A
Страница 66: ...info_cl fm...