ToxiRAE Pro User’s Guide
39
12.
In the menu that now appears on the left side, click “Policy Enforcement.” It is
highlighted, and the Policy Enforcement pane is shown:
For “Must Calibrate” and “Must Bump,” you have the options of no enforcement or
enforcement (including “Can’t Bypass,” and “Can Bypass”).
Must Calibrate.
The user is prompted to calibrate the instrument when calibration is due
(as set by the calibration interval). There are two programmable options:
•
Can’t Bypass.
Unless calibration is performed, the instrument cannot be used,
and the only option is to turn off the instrument.
•
Can Bypass.
If calibration is due but the user does not want to perform a
calibration, the instrument can still be used.
Must Bump
•
Can’t Bypass.
Unless a bump test is performed, the instrument cannot be used,
and the only option is to turn off the instrument.
•
Can Bypass.
If a bump test is due but the user does not want to perform one, the
instrument can still be used.