ToxiRAE Pro User’s Guide
14
4.
After self-testing, two start-up options are available for the ToxiRAE Pro: Normal
and Fast Start-up. The following information is displayed in sequence under Normal
Start-up and skipped under Fast Start-up:
•
Sensor type
•
Sensor serial number
•
Date, time, and temperature
•
Calibration gas
•
Last calibration date
•
Measurement gas
•
User mode
•
Battery type, voltage, and shutoff voltage
•
Alarm mode
•
Buzzer enabled/disabled status
•
Light enabled/disabled status
•
Vibration alarm enabled/disabled status
•
Alarm settings for High, Low, STEL, and TWA (as applicable)
•
Datalog type and interval
5.
Once the startup process is complete, a sensor warm-up screen is shown (with
countdown, if necessary; the countdown time may vary for different sensors). Then
the main measurement screen is shown. The type of monitored hazard is displayed,
with the current reading. The ToxiRAE Pro is now ready for use.
8. Turning The ToxiRAE Pro Off
Press and hold [MODE]. In 2 seconds, a 5-second countdown to shutoff begins,
accompanied by one buzzer beep per second. You must hold your finger on the key for
the entire shutoff process. If you remove your finger from the key during the countdown,
the shutoff operation is canceled and the ToxiRAE Pro continues normal operation.
When you see “Unit Off,” release your finger from the [MODE] key. The ToxiRAE Pro
is now off.
9. Testing The Alarms (Manually)
Under normal non-alarm conditions, the buzzer, vibration alarm, and LEDs can be tested
at any time by pressing and holding [Y/+].
Important!
If any of the alarms does not respond to this test, check the Alarm Settings. It is
possible that any or all of the alarms have been turned off. If all the alarms are turned on, but
one or more of them (buzzer, LED lights, or vibration alarm) does not respond to this test, do
not use the instrument; contact your RAE Systems distributor for technical support.