12.4. Concentration [Value]
You can set the concentration by double-clicking in the respective gas concentration box
and then typing in the concentration value.
12.5. Concentration Unit
When you pull down the Concentration Unit menu, select the desired gas concentration
units (there are other types of units).
ppm
%
ppb
mg
ug
%LEL
%VOL
%CH
4
12.6. Purge Time (Sec.)
Type to set the number of seconds for the system to purge with fresh air after performing
a bump test or calibration.
12.7. Soak Time (Sec.)
Type to set the number of seconds for the system to allow the sensor to be pre-exposed to
calibration gas before bumping or calibrating.
Содержание AutoRAE 2
Страница 1: ......
Страница 38: ...Select Exit to return the monitor to Normal Measurement Mode...
Страница 43: ...When you click on Gas Config 8 the two gas bottles Gas Inlet 15 and Gas Inlet 16 are shown...
Страница 76: ......
Страница 95: ...Set the password for access to the AutoRAE 2 Note The default value is 0000...
Страница 106: ...6 This screen appears 7 Expand either Online or Offline to view Instruments 8 Click Instruments...
Страница 107: ...Instruments that have been bump tested and calibrated on a system with this AutoRAE 2 Controller are shown...
Страница 108: ...9 Double click on an instrument to view its reports The Reports window opens for this instrument...
Страница 121: ......