22. Updating Firmware On The AutoRAE 2 Controller
Updates to the AutoRAE 2 Controller’s firmware may be produced, and these can be
loaded into the AutoRAE 2 Controller using ProRAE Studio II software running on a PC.
1.
Download firmware from the RAE Systems web site or from a CD-ROM.
2.
Connect a PC running ProRAE Studio II to the AutoRAE 2 Controller via a USB
cable.
3.
Start ProRAE Studio II.
4.
Click “Administrator.”
5.
Enter the password (the default is “rae”).
6.
Click “OK.”
7.
Click “Detect the instruments automatically.”