23.2. Start mass adjustment
It is possible to adjust only a start mass, it helps to correct the start zero
when the span does not change.
Procedure:
•
Enter
<P10 USER CALIB.> according to 11.2. and then:
•
Unload the scale,
•
Press
. During adjusting a start mass you will see:
ADJ. START MASS, in the bottom line,
•
After completing this procedure the scale will return to the
following window:
•
Return to weighing, saving parameters.
Return to weighing:
See 11.3.
- 97 -