- 158 -
28.4.4. Set tare value
Syntax: UT_TARE CR LF, where TARE – tare value
Possible answers:
UT_OK CR LF
- command executed
UT_I CR LF
- command comprehended but cannot be executed
ES CR LF
- command not recognised (possible wrong tare format)
Notice:
This protocole uses the dot character as a decimal point for separating
the decimal fraction part.
28.4.5. Send the stable result in basic unit
Syntax: S CR LF
Possible answers:
S_A CR LF
S_E CR LF
- command accepted and in progress
- time limit for stable result exceeded
S_I CR LF
- command comprehended but cannot be executed
S_A CR LF
MASS FRAME
- command accepted and in progress
- mass value in basic unit is returned
Frame format:
1 2-3
4
5 6 7-15 16 17
18
19
20
21
S space stability space sign
mass
space unit CR
LF
Example:
S CR LF – computer command
S _ A CR LF - command accepted and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF – command done, mass value
in basic unit is returned.