- 145 -
27.5.1. Technical specification
Parameters of outputs
Quantity of outputs
8
Type of outputs
Reed operation contacts
Wire diameter
0,14 - 0,5mm
2
Maximal load-current contact capacity
0,2A DC
Maximal forward voltage
50V DC
Parameters of inputs
Quantity of inputs
8
Input type
Optoinsulated
Wire diameter
0,14 – 0,5mm
2
Control voltage range
5 -24V DC
27.5.2. Installing method in PUE C41H indicators
Module is designated for assembly inside indicator PUE C41H. Module
is assembled to main board of indicator to 10-pin J12 connector. For
module WE8 gland is installed on casing lid, 3m cable ended with
insulated conductors is led through gland.
Installing procedure:
1. Unplug the indicator (remove the plug from socket 230V);
2. Unscrew the casing lid (back part of indicator casing);
3. Install the module in connector J12 on main board;
4. While installing module pay attention to plastic posts fastening
to main board. They should be placed in assembly holes in main
board and in assembly holes in module WE 8,
5. In one of free glands remove the stopper and led through it cable
IN/OUT (if necessary use bigger gland depending on cable diameter);
6. Connect the cable IN/OUT to joint J1, J2 for outputs and J3, J4 for
inputs on module 8IN/8OUT according to description given in table;
7. Connect cable IN/OUT to group of conductors (undo band clips
fastening group of conductors, lay the cable and clamp band clips –
multiple use band clips);
8. Assembly cover of indicator casing.
Содержание WPW/E
Страница 2: ...2 NOVEMBER 2011...
Страница 129: ...Scale printer CITIZEN EPSON cable Scale Zebra printer cable Scale 3IN 3OUT RSTS 8 184 2M cable 129...
Страница 146: ...A WE 8 module placement on the main board of PUE C41H 27 5 3 I O diagram WE8 inputs diagram WE8 outputs diagram 146...
Страница 185: ...32 6 Examples of surge protections Connection diagram of dosing indicators outputs with protection for DC 185...
Страница 186: ...Connection diagram of dosing indicators outputs with protection for AC 186...