Mass - 9 characters justified to the right
Unit
- 3 characters justified to the left
28.4.22. Send all implemented commands
Syntax: PC CR LF
Possible answers:
PC_A_”Z,T,S,SI,SU,SUI,C1,C0,CU1,CU0,PC,K1,K0,DH,UH,
ODH,OUH,S1,S0,OT,UT" – command executed, the indicator
have sent all the implemented commands.
28.5. Manual printouts / automatic printouts
Users can general manual or automatic printouts from the scale.
•
Manual printouts can be performed after loading the pan and
stabilizing indication by pressing
.
•
Automatic printouts can be performed only after loading the
pan and stabilizing indication.
Notice:
If a scale is verified printouts of immediate values are blocked.
Format frame:
1
2
3 4
-12 13 14 15 16 17 18
stability space
sign
mass space
unit
CR LF
Stability character
[space] if stable
[?] if not stable
[^] if an indication over the range
[v] if fan indication below the range
sign
[space] for positive values or
[-] for negative values
mass
9 characters justified to the right
unit
3 characters justified to the left
command
3 characters justified to the left
- 165 -
Содержание WPW/E
Страница 2: ...2 NOVEMBER 2011...
Страница 129: ...Scale printer CITIZEN EPSON cable Scale Zebra printer cable Scale 3IN 3OUT RSTS 8 184 2M cable 129...
Страница 146: ...A WE 8 module placement on the main board of PUE C41H 27 5 3 I O diagram WE8 inputs diagram WE8 outputs diagram 146...
Страница 185: ...32 6 Examples of surge protections Connection diagram of dosing indicators outputs with protection for DC 185...
Страница 186: ...Connection diagram of dosing indicators outputs with protection for AC 186...