29. ERROR MESSAGES
ERROR ”XXX” ESC – RETURN TO PREVIOUS SETTING
(where: XXX – parameter name) – confirmed with ENTER
of a wrong value in the user’s menu,
Err2
- Value beyond the zero range,
Err3
- Value beyond the tare range,
Err4
- Calibration mass or start mass adjustment error
(
±
1% for weight,
±
10% for start mass),
Err5
- Mass of a single piece lower than the 0.1 of scale division,
Err6
- Mass of all pcs (declared as standard) lower than one
division,
Err8
- Tarring / zeroing operation time exceeded,
NULL
- Zero value from the AD converter,
FULL2
- Measurement
range overflow,
HI
- Display range overflow,
LH
- Start mass error, the mass on the weighing platform
is beyond the acceptable range
±
20 of start mass
30. TROUBLE SHOOTING
Problem Cause
Solution
Scale switches off
automatically
„t1” parameter set to „YES”
(Power save)
In <P9 OTHERS> change
<P9.3 AUTO SWITCH.> to
„NO”
„LH” appears on the
display
Weight pan not empty
when switching on
Unload the pan. Indication
returns to zero