Connect the scale to a computer/switch using a PT0017 or PT0014
cable (depending on the Ethernet module on board – see ch. 27.1.2),
•
Set the port for communication with a computer – parameter
<P3.1.1 COMP. PORT> set to Ethernet (see ch. 15.1.1),
•
Set parameters from the <P2.4 ETHERNET> group (see ch. 14.2):
−
Parameter
<P2.4.1 WORK MODE> should be set to SERVER,
−
Parameters
<P2.4.2 IP ADDRESS>, <P2.4.3 SUBNET MASK>,
<P2.4.4 GATEWAY> should be set according to the network
supervisor directions,
−
<P2.4.5 LOCAL PORT> set the same as in the computer
program (in networks some ports can be taken),
−
Parameter
<P2.4.8 TIMEOUT> set to 0 in order to prevent
breaking connection.
•
Press
the
„Settings” button:
•
In the overlap <Transmission setting>:
−
Set the communication type „Ethernet Transmission TCP”,
−
„IP address 1” the same as <P2.4.2 IP ADDRESS> in the scale,
−
„Port”/”Com” the same as <P2.4.5 LOCAL PORT> in the scale,
•
Confirm the changes by clicking
,
•
Rerun the program,
•
Press
the
„Connect” button,
•
The appropriate communication with scales is signalled by pictogram
in the left part of the display,
- 174 -