- 157 -
Z_A CR LF
Z_^ CR LF
- command accepted and in progress
- command comprehended but zero range overflow appeared
Z_A CR LF
Z_E CR LF
- command accepted and in progress
- time limit for stable result exceeded
Z_I CR LF
- command comprehended but cannot be executed
28.4.2. Tarring
Syntax: T CR LF
Possible answers:
T_A CR LF
T_D CR LF
- command accepted and in progress
- command completed
T_A CR LF
T_v CR LF
- command accepted and in progress
- command comprehended but tare range overflow appeared
T_A CR LF
T_E CR LF
- command accepted and in progress
- time limit for stable result exceeded
T_I CR LF
- command comprehended but cannot be executed
28.4.3. Get tare value
Syntax: OT CR LF
Possible answers:
OT_TARA
CR LF
- command executed
Frame format:
1 2
3
4-12
13 14 15 16
17
18 19
O T space
tare
space unit space
CR
LF
Tare - 9 characters with decimal point justified to the right
Unit
- 3 characters justified to the left
Содержание WPW/E
Страница 2: ...2 NOVEMBER 2011...
Страница 129: ...Scale printer CITIZEN EPSON cable Scale Zebra printer cable Scale 3IN 3OUT RSTS 8 184 2M cable 129...
Страница 146: ...A WE 8 module placement on the main board of PUE C41H 27 5 3 I O diagram WE8 inputs diagram WE8 outputs diagram 146...
Страница 185: ...32 6 Examples of surge protections Connection diagram of dosing indicators outputs with protection for DC 185...
Страница 186: ...Connection diagram of dosing indicators outputs with protection for AC 186...