- 171 -
5. A special protocol is required in order the code be received by
RADWAG equipment. It is required to program an appropriate
prefix and suffix. Prefix – one byte 01 hexadecimally, suffix one
byte 0D hexadecimally.
6. Most barcode scanners allow to enable/disable different
symbologies (barcode types).
7. Programming of scanners is usually performed by reading special
barcodes or by using an external software tool.
8. Scanners marketed together with RADWAG systems are always
configured according to the rules above.
Barcode with required prefix and
suffix in hexadecimal format
Barcode without
required –fixes in
ASCII format
Code type
01 30 30 32 31 30 31 32 36 0D
00210126
EAN-8
01 30 31 32 33 34 35 36 37 38 39 0D
0123456789
2 of 5
01 43 4F 44 45 20 33 39 20 54 45 53 54 0D
CODE 39 TEST
CODE 39
01 31 31 30 31 32 33 34 35 36 37 38 39 31 0D
1101234567891
EAN-13
01 43 6F 64 65 20 31 32 38 20 54 65 73 74 0D
CODE 128 Test
CODE 128
32.2. Cooperation with „EDYTOR WPW” – PC software tool
Standard WPW scales can be supervised by the accompanied program
EDYTOR WPW, that can work as an additional display on the monitor and:
•
Edit and change user parameters (filters, date/time, backlight,
configuring I/O, port parameters etc.),
•
Create, upload, edit, download databases,
•
Receive and save printouts in text files,
•
Define the bottom line,
•
Define non-standard printouts.
Содержание WPW/E
Страница 2: ...2 NOVEMBER 2011...
Страница 129: ...Scale printer CITIZEN EPSON cable Scale Zebra printer cable Scale 3IN 3OUT RSTS 8 184 2M cable 129...
Страница 146: ...A WE 8 module placement on the main board of PUE C41H 27 5 3 I O diagram WE8 inputs diagram WE8 outputs diagram 146...
Страница 185: ...32 6 Examples of surge protections Connection diagram of dosing indicators outputs with protection for DC 185...
Страница 186: ...Connection diagram of dosing indicators outputs with protection for AC 186...